Vis-U-All High Temp 6"x10" Heat Seal Pouch 200 pouches per box; 5 boxes per case
Steris Corporation began a recall of Vis-U-All High Temp 6"x10" Heat Seal Pouch 200 pouches per box; 5 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable... on . FDA classified it as Class II and gives the reason as "On June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0757-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- STERIS Corporation
- Product
- Vis-U-All High Temp 6"x10" Heat Seal Pouch 200 pouches per box; 5 boxes per case Product Usage: The Vis-U-All High Temperature Sterilization Pouch is suitable for use by healthcare providers to enclose and seal medical devices to be sterilized by steam. The pouch is designed to maintain sterility of properly processed medical devices during normal handling and storage until the pouch is opened and the medical device is removed for use.
- Product code
- FRG Wrap, Sterilization
- Reason for recall
- On June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.
- Root cause tag
- labeling and UDI
- FDA root cause
- Component change control
- Action
- Steris sent an Urgent Voluntary Recall Notice letter dated July 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately inspect on-hand inventory for recalled product. If you have product in your possession, please quarantine it and contact firm to coordinate shipment of replacement product. If you have questions regarding this matter, please contact STERIS Customer Service at 1-800-548-4873.
- Quantity in commerce
- 55
- Distribution
- US Nationwide Distribution
- Product codes and lots
- Model# 883610 Lot# 161124A and #170224A
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0743-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0744-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0745-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0746-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0747-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0748-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0749-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0750-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0751-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0752-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0753-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0754-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0755-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0756-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0757-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0758-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0759-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0760-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0761-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0762-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0763-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0764-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0765-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0766-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0767-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0768-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0769-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0770-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0771-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.