Z-0776-2018Class II

Intellivue iX Information Center Software - All PIIC iX Surveillance stations including: 866023 IntelliVue...

Philips North America LLC / initiated 2018-01-03 / Terminated / root cause: software

Philips Electronics North America Corporation began a recall of Intellivue iX Information Center Software - All PIIC iX Surveillance stations including: 866023 IntelliVue Info Center iX A.0 866117 PIIC Classic Upgrade... on . FDA classified it as Class II and gives the reason as "Once a surveillance station is restarted on January 1, 2018 or later, the station will be unable to perform patient discharge and transfer operations. Any subsequent attempt to perform these operations will cause the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0776-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Intellivue iX Information Center Software - All PIIC iX Surveillance stations including: 866023 IntelliVue Info Center iX A.0 866117 PIIC Classic Upgrade 866389 IntelliVue Info Center iX B.0, C.0 867141 IntelliVue Info Center iX B.0 866025 IntelliVue Server IX A.0 866118 Database Server Upgrade A.0
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
Once a surveillance station is restarted on January 1, 2018 or later, the station will be unable to perform patient discharge and transfer operations. Any subsequent attempt to perform these operations will cause the station to restart, resulting in a short period of loss of monitoring at the Surveillance station during such restart. Until this issue can be corrected, users should avoid intentionally restarting their Surveillance stations in 2018.
Root cause tag
software
FDA root cause
Software design
Action
To avoid this issue, do not intentionally restart any surveillance stations that have been running normally since the new year. You may use the workflows described in attachment Alternative Workflows for PIIC iX to avoid operations that may cause the surveillance station to restart. When a patient is discharged, confirm that the patient is discharged. Philips has initiated a correction to address this issue and will provide this software update to customers with affected devices at no charge. Software updates for PIIC iX A.0x and B.0x are expected to be available by January 8th. A Philips Healthcare representative will contact customers with affected devices to arrange for correction of the issue. Contact your local Philips representative if you have a question about any device affected by this issue.
Quantity in commerce
20357
Distribution
Domestic and Foreign
Product codes and lots
All PIIC iX Surveillance stations including: 866023 IntelliVue Info Center iX A.0 866117 PIIC Classic Upgrade 866389 IntelliVue Info Center iX B.0, C.0 867141 IntelliVue Info Center iX B.0 866025 IntelliVue Server IX A.0 866118 Database Server Upgrade A.0

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0776-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0776-2018Class IITerminatedVoluntary: Firm initiated