Z-0847-2018Class II

A-Series A7 Anesthesia Delivery System; Model Number: 0632F-PA0000X (A7)

Datascope Corp. / initiated 2017-07-14 / Terminated / root cause: design

Mindray DS USA, Inc. dba Mindray North America began a recall of A-Series A7 Anesthesia Delivery System; Model Number: 0632F-PA0000X (A7) Product Usage: The A-Series Anesthesia System is a device used to administer to a... on . FDA classified it as Class II and gives the reason as "A software issue may result in the previous settings being applied instead of the default settings or the unit may skip the startup leak test". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0847-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Datascope Corp.
Product
A-Series A7 Anesthesia Delivery System; Model Number: 0632F-PA0000X (A7) Product Usage: The A-Series Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic, and to maintain a patient s ventilation. The A-Series is intended for use by licensed clinicians, for patients requiring anesthesia within a health care facility, and can be used for adult, pediatric, and infant populations.
Product code
BSZ Gas-Machine, Anesthesia
Reason for recall
A software issue may result in the previous settings being applied instead of the default settings or the unit may skip the startup leak test.
Root cause tag
design
FDA root cause
Device Design
Action
Mindray sent an Urgent Medical Device Correction letter dated June 9, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. Corrective Action: In order to eliminate any potential for the EEPROM issue to manifest, Mindray is offering two options for upgrading affected systems in your facility: Option 1: You may perform the software upgrade if your facility has Biomedical Engineering or other staff that have been trained by Mindray to perform maintenance operations on the ASeries. Please contact Mindray's Technical Support team who will assist with providing the software and instructions for performing the software upgrade. The Technical Support can be reached at (877) 913-9663 (Option 1). Option 2: A Mindray Service Representative will upgrade the affected systems in your facility at no charge. Please contact your Mindray Service Representative to arrange for this upgrade at (877) 913-9663 (Option 8). Both the Technical Support and Service teams may be reached Monday through Friday, 8:30am - 5:30pm, ET.
Quantity in commerce
2875 in total
Distribution
Worldwide Distribution - US Nationwide
Product codes and lots
Units distributed or upgraded between March 2013 and December 21, 2016. Serial numbers requested.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0847-2018