Z-0840-2018Class II

Brilliance Big Bore Oncology CTwith 4.2.0 software version. Model #' 728243

Philips North America LLC / initiated 2018-01-03 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of Brilliance Big Bore Oncology CTwith 4.2.0 software version. Model #' 728243 Product Usage: Computed Tomography X-ray systems intended to produce... on . FDA classified it as Class II and gives the reason as "Software issues including: Software errors that may result in Extended Field of View check message not displayed when Field of View is set >600 mm for a paused scan; Request for change to automatically turn on the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0840-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Brilliance Big Bore Oncology CTwith 4.2.0 software version. Model #' 728243 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Software issues including: Software errors that may result in Extended Field of View check message not displayed when Field of View is set >600 mm for a paused scan; Request for change to automatically turn on the center x/y 0 setting and the disable result rotation setting when turning on the CT simulation Exam Card; Software errors that may result in Tumor LOC allowing POI (Point of Interest) locking when a non-isocenter slice is displayed; Software errors that may result in partial set of images or inability to generate CT images.
Root cause tag
software
FDA root cause
Under Investigation by firm
Action
Philips sent an Urgent Field Safety Notice Medical Device Correction letter dated January 2, 2018 to their customers. The letter identified the affected product problem and actions to be taken. For questions contact your local Philips representative or local Philips Healthcare office.
Quantity in commerce
11
Distribution
Worldwide Distribution - US Nationwide
Product codes and lots
Brilliance Big Bore Oncology CT, Model # 728243 Serial #s: 76025 76022 76024 76023 76016 76021 76008 76010 76002 76020 76011

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0840-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0839-2018Class IITerminatedVoluntary: Firm initiated
Z-0840-2018Class IITerminatedVoluntary: Firm initiated