Z-0775-2018Class II
979195 Loaner Kit for femoral intramedullary nail. Orthopedic instrument
Zimmer GmbH / initiated 2017-09-13 / Terminated / root cause: design
Zimmer Biomet, Inc. began a recall of 979195 Loaner Kit for femoral intramedullary nail. Orthopedic instrument on . FDA classified it as Class II and gives the reason as "Fracture of the bolt connecting the insertion guide to the intramedullary nail during insertion of the femoral nail". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0775-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- 979195 Loaner Kit for femoral intramedullary nail. Orthopedic instrument.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Fracture of the bolt connecting the insertion guide to the intramedullary nail during insertion of the femoral nail.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On September 13, 2017 an URGENT MEDICAL DEVICE RECALL REMOVAL was issued to customers requesting that they review inventory and quarantine affected product, which will be immediately returned to recalling firm. Also to notify any customers who may have received product. Questions or concerns can be directed to Customer Service at 574-371-3071
- Quantity in commerce
- 337 total
- Distribution
- Distribution includes US Nationwide.
- Product codes and lots
- Finished Lot order: M370490 M354450 M321540 M975460 M639280 M426760 M629950 M629930 M073530 M073540 M073550 M073560 M073570 M073580 M073590 M073600 M073610 M627490 M656470 M656580 M656590 M656600 M656610 M656620 M656640 M656660 M966880 M967050 M967060 M967070 M980260 M980270 M508490 M508410 M508500 M508450 M508480 M508520 M508470 M508510 M508460
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0773-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0774-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0775-2018 | Class II | Terminated | Voluntary: Firm initiated |