Z-0810-2018Class II

AccuLIF TL Insertion Handle The AccuLIF Insertion Handle attaches to the implant and is specifically designed...

Howmedica Osteonics Corp / initiated 2017-09-08 / Terminated

Howmedica Osteonics Corp. began a recall of AccuLIF TL Insertion Handle The AccuLIF Insertion Handle attaches to the implant and is specifically designed to allow a surgeon to insert the implant into the... on . FDA classified it as Class II and gives the reason as "When the lever of the Insertion Handle is in the "locked" position, the implant may be able to pivot rather than maintaining a rigid connection". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0810-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
AccuLIF TL Insertion Handle The AccuLIF Insertion Handle attaches to the implant and is specifically designed to allow a surgeon to insert the implant into the disc space with a fixed engagement between the two components to permit a controlled positioning of the implant in the anterior portion of the intervertebral space, and then through activation of the rotational pivot between the Handle and implant for final placement of the crescent-shaped cage across the anterior portion of the disc space. The AccuLIF TL Cages are indicated for intervertebral body fusion with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion in patients with degenerative disc disease (DOD) at one level or two contiguous levels from L2 to S1.
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
When the lever of the Insertion Handle is in the "locked" position, the implant may be able to pivot rather than maintaining a rigid connection.
FDA root cause
Under Investigation by firm
Action
A recall letter, dated August 25, 2017, titled "Product Recall Re: AccuLIFTL Insertion Handle" instructing customers to meet with the Stryker Representative issuing the letter, complete the requested form and return any affected product to this representative. Questions or concerns can be directed to Customer Service at 201-749-8699
Quantity in commerce
130
Distribution
Nationally
Product codes and lots
Model# 800804 Lot numbers 160010 and 170003

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0810-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0810-2018Class IITerminatedVoluntary: Firm initiated