Z-0841-2018Class II

IQon Spectral CT with 4.7.2 software version.- Model 728332, a Philips Computed Tomography X-Ray System

Philips North America LLC / initiated 2018-01-11 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of IQon Spectral CT with 4.7.2 software version.- Model 728332, a Philips Computed Tomography X-Ray System on . FDA classified it as Class II and gives the reason as "23 Software related issues for the IQon Spectral CT with 4.7.2 software version. These issues include: " Software errors that may result in CT spectral images misrepresentation or incorrect treatment; " Software errors...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0841-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
IQon Spectral CT with 4.7.2 software version.- Model 728332, a Philips Computed Tomography X-Ray System
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
23 Software related issues for the IQon Spectral CT with 4.7.2 software version. These issues include: " Software errors that may result in CT spectral images misrepresentation or incorrect treatment; " Software errors that may result in partial set of images or inability to generate CT images.
Root cause tag
software
FDA root cause
Under Investigation by firm
Action
Philips Healthcare is notifying the affected users of these issues via a Field Safety Notice and will be releasing a software fix via Field Change Order to correct these issues. They also provide the following instructions.: Issue 1, Uric acid application for gout provides inconsistent results - When reviewing images using the Uric Acid Spectral Result type inconsistent results were provided on gout studies. The tendons would occasionally be displayed as containing Uric Acid even with no or normal uric acid blood levels. Issue 2, Multi-phase Cardiac CTA Scans should include a planned conventional result for every planned spectral result - When planning a Cardiac multi-phase study, it is possible to create a phase that consists of only a Spectral Result. Issue 3, 0-MAR information is missing in DICOM when saving as original or Secondary Capture - When an SBl that was created with 0-MAR enabled is used to create and save any Spectral result: " DICOM information about 0-MAR is missing but the 0-MAR label appears on a result saved as Original " DICOM information and the 0-MAR label are both missing on a result saved as Secondary Capture Issue 4, Attenuation plot presents incorrect HU values for MonoE - When comparing HU values of an ROl to the values in the attenuation plot graph when using the Show spectral plots for this ROI option, there is a difference of 10-20 HU values between the ROI HU value and the value shown on the graph Issue 5, Windows Operating System Time Zone Updates - Microsoft Windows Daylight Savings Time Zone updates have not been applied since December 2014. This may cause the system time to be incorrect at the start and end of daylight savings time. Issue 6, Sending Calcium Score findings to report using Mass Score protocol provides confusing information - The Heartbeat CS (Calcium Scoring) option provides confusing information in the Impressions section of the Calcium Scoring report. The report provides m
Quantity in commerce
63
Distribution
Domestic and Foreign
Product codes and lots
Serial Numbers: 860008 860009 860010 860011 860012 860013 860014 860015 860016 860017 860018 860019 860020 860021 860022 860023 860024 860025 860026 860027 860028 860029 860030 860031 860032 860033 860034 860035 860036 860037 860038 860039 860040 860041 860042 860043 860046 860048 860049 860050 860051 860052 860053 860054 860055 860056 860057 860058 860059 860060 860061 860062 860063 860064 860065 860066 860067 860068 860070 860071 860072 890002 860047

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0841-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0841-2018Class IITerminatedVoluntary: Firm initiated