Z-0708-2018Class II
Axiom Artis, Model Numbers: 5904433, 5904441, 5904649, 5904656, 5917054, 7007755, 7008605, 7412807, 7413078...
Siemens Medical Solutions USA, Inc. / initiated 2017-06-06 / Terminated / root cause: design
Siemens Medical Solutions USA, Inc began a recall of Axiom Artis, Model Numbers: 5904433, 5904441, 5904649, 5904656, 5917054, 7007755, 7008605, 7412807, 7413078, 7555357, 7555373, 7727717, 7728350 & 7728392 on . FDA classified it as Class II and gives the reason as "Potential for liquid entry". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0708-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Axiom Artis, Model Numbers: 5904433, 5904441, 5904649, 5904656, 5917054, 7007755, 7008605, 7412807, 7413078, 7555357, 7555373, 7727717, 7728350 & 7728392
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Potential for liquid entry
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On June 6, 2017, Siemens Healthineers distributed URGENT MEDICAL DEVICE CORRECTION notices to their customers via USPS Certified Mail. Siemens conducted a corrective field action in June 2015 to protect the Artis table against liquid entry at critical locations. However, even with this corrective action, if there are larger amounts of liquid on the table surface, they can still enter the table. Siemens Medical Solutions USA, Inc. will contact customers to arrange a date to perform the corrective action. Please feel free to contact the service organization for an earlier appointment by calling 610-448-6461.
- Distribution
- U.S. Nationwide
- Product codes and lots
- 10094135, 10094137, 10094141, 10094142, 10094143, 10280959, 10848280, 10848281, 10848282, 10848353, 10848354, 10848355, 5904433, 5904441, 5904649, 5904656, 5917054, 7007755, 7008605, 7412807, 7413078, 7555357, 7555373, 7727717, 7728350, 7728392
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0706-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0707-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0708-2018 | Class II | Terminated | Voluntary: Firm initiated |