510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0673-2018Class II

Vanguard Complete Knee System, Vanguard CR Tibial Bearing 12x71/75, Item Number: 183442

Zimmer GmbH / initiated 2017-09-29 / Terminated / root cause: labeling and UDI

Zimmer Biomet, Inc. began a recall of Vanguard Complete Knee System, Vanguard CR Tibial Bearing 12x71/75, Item Number: 183442 Product Usage: For use in total knee arthroplasty on . FDA classified it as Class II and gives the reason as "Knee implant components (tibial bearings) are labeled with the incorrect size". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0673-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Vanguard Complete Knee System, Vanguard CR Tibial Bearing 12x71/75, Item Number: 183442 Product Usage: For use in total knee arthroplasty
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Knee implant components (tibial bearings) are labeled with the incorrect size.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
Zimmer Biomet sent an Urgent Medical Device Recall letter dated September 29, 2017 to their customers. The letter identified the affected product, problems and actions to be taken. Customers were instructed to remove the affected product from inventory and provide to their Zimmer sales representative or, if the consignee is a distributor, to ship the affected product back to Zimmer. Consignees were also asked to complete a recall response form and send this back to the firm. Returns are being physically destroyed. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected].
Quantity in commerce
24
Distribution
Worldwide Distribution - US Nationwide in the states of CA, GA, KS, LA, MA, MI, MN, OH, TX, VA, WA, WI, and the countries of Netherlands
Product codes and lots
Lot Number: 388680, Exp. Date 5/19/2022

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0673-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0673-2018Class IITerminatedVoluntary: Firm initiated
Z-0674-2018Class IITerminatedVoluntary: Firm initiated