Z-0895-2018Class II
Hemosll ReadiPlasTin (20 ml Size), Part Number 0020301400
Instrumentation Laboratory Co. / initiated 2017-08-02 / Terminated
Instrumentation Laboratory Co. began a recall of Hemosll ReadiPlasTin (20 ml Size), Part Number 0020301400 on . FDA classified it as Class II and gives the reason as "Instrumentation Laboratory Co. received customer reports of performance issues with some vials of Lot Nos. N1166235 and N0177760, including increased imprecision, out of range quality controls and prolonged sample...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0895-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Instrumentation Laboratory Co.
- Product
- Hemosll ReadiPlasTin (20 ml Size), Part Number 0020301400
- Product code
- GJS Test, Time, Prothrombin
- Reason for recall
- Instrumentation Laboratory Co. received customer reports of performance issues with some vials of Lot Nos. N1166235 and N0177760, including increased imprecision, out of range quality controls and prolonged sample results.
- FDA root cause
- Under Investigation by firm
- Action
- On August 2, 2017, Instrumentation Laboratory sent an Urgent Medical Device Removal letters dated August 2, 2017 to their customers. On October 5, 2017 Instrumentation Laboratory expanded their recall adding 3 additional lots to a second Urgent Medical Device Removal letter. On February 26, 2018, Instrumentation Laboratory expanded their recall a 3rd time to include 1 additional lot to a third Urgent Medical Device Removal letter. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete a Mandatory Tracking Response form. For questions contact your Customer Service Representative at 1-800-955-9525, option #1.
- Quantity in commerce
- 14,894 kits
- Distribution
- Worldwide Distribution - US Nationwide
- Product codes and lots
- Lots: N1166235, N0177760, N0278358, N0478057, N0479504 & N0278177.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0895-2018 | Class II | Terminated | Voluntary: Firm initiated |