510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0908-2018Class II

Endo GIA" Radial Reload with Tri-Staple" Technology The Endo GIA radial reloads with Tri-Staple Technology...

Medtronic, Inc. / initiated 2017-09-15 / Terminated / root cause: manufacturing process

Covidien LLC began a recall of Endo GIA" Radial Reload with Tri-Staple" Technology The Endo GIA radial reloads with Tri-Staple Technology have application in open or minimally invasive... on . FDA classified it as Class II and gives the reason as "The device cartridge disengaged during use due to manufacturing error". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0908-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Endo GIA" Radial Reload with Tri-Staple" Technology The Endo GIA radial reloads with Tri-Staple Technology have application in open or minimally invasive general abdominal, gynecologic and thoracic surgery for resection and transection of tissue and creation of anastomosis, as well as application deep in the pelvis, e.g. low anterior resection. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas.
Product code
GDW Staple, Implantable
Reason for recall
The device cartridge disengaged during use due to manufacturing error.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Medtronic sent an "URGENT MEDICAL RECALL LETTER" dated September 13, 2017, was issued to customers titled "Covidien Endo GIA Black Radial Reload with Tri-Staple Technology" urging customers to quarantine and return unused product to recalling firm. Questions or concerns can be directed to: [email protected]. For further questions, please call (203) 492-5000.
Quantity in commerce
163
Distribution
Internationally, including Japan. No USA Customers
Product codes and lots
Product number: EGIARADXT Lot code: N6L0351X

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0908-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0908-2018Class IITerminatedVoluntary: Firm initiated