510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0636-2018Class II

Endo GIA Auto Suture Universal Articulating Loading Unit for use in abdominal, gynecologic, pediatric and...

Medtronic, Inc. / initiated 2017-09-13 / Terminated

Covidien LLC began a recall of Endo GIA Auto Suture Universal Articulating Loading Unit for use in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and... on . FDA classified it as Class II and gives the reason as "Potential for failure of the device safety interlock. The safety interlock prevents an empty single-use loading unit from being fired a second time". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0636-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Endo GIA Auto Suture Universal Articulating Loading Unit for use in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis.
Product code
GDW Staple, Implantable
Reason for recall
Potential for failure of the device safety interlock. The safety interlock prevents an empty single-use loading unit from being fired a second time.
FDA root cause
Under Investigation by firm
Action
All consignees were notified via Federal Express or certified mail on September 13, 2017, and the letter informs customers of the potential for failure of the device safety interlock and the actions they should take. Customers are requested to acknowledge receipt and understanding of the Urgent Field Corrective Action Notice. All in control inventory and product returned from customers will be scrapped.
Quantity in commerce
3,816
Distribution
Nationally
Product codes and lots
N7F1058KX, N7F1121KX, N7F1056KX, N7F1057KX

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0636-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0636-2018Class IITerminatedVoluntary: Firm initiated