Z-0694-2018Class II

Footswitch accessory to the CUSA Clarity Ultrasonic Tissue Ablation System; Catalogue No: C7000

Integra LifeSciences Corporation / initiated 2017-09-26 / Terminated / root cause: manufacturing process

Integra LifeSciences Corp. began a recall of Footswitch accessory to the CUSA Clarity Ultrasonic Tissue Ablation System; Catalogue No: C7000 Product Usage: The footswitch is a component of the CUSA... on . FDA classified it as Class II and gives the reason as "A faulty footswitch may result in unintended ultrasonic fragmentation activation. There are two scenarios when the failure mode may occur: 1) Ultrasonic fragmentation may be inadvertently activated at start-up or at the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0694-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra LifeSciences Corporation
Product
Footswitch accessory to the CUSA Clarity Ultrasonic Tissue Ablation System; Catalogue No: C7000 Product Usage: The footswitch is a component of the CUSA Clarity Ultrasonic Surgical Aspirator System, which is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft tissue is desirable. The CUSA Clarity Ultrasonic Surgical Aspirator is indicated for use in: Neurosurgery, Plastic and Reconstructive surgery, Orthopedic Surgery, Gynecological Surgery and Thoracic Surgery and the following specific uses are cleared for use in the US only: Gastrointestinal and Affiliated Organ Surgery including removal of benign or malignant tumors or other unwanted tissue, including hepatic parenchyma, in open or laparoscopic procedures, hepatic resection, tumor resection, lobectomy or trisegmentectomy, or removal of tissue during liver allotransplantation and donor hepatectomy Urological surgery including removal of renal parenchyma during nephrectomy or partial nephrectomy General Surgery including removal of benign or malignant tumors or other unwanted soft tissue in open or minimally invasive general surgical procedures Laparoscopic Surgery including removal of hepatic parenchyma in laparoscopic hepatic resection, lobectomy or trisegmentectomy, in laparoscopic donor hepatectomy or laparoscopic cholecystectomy or laparoscopic pancreatic jejunostomy, or pancreatectomy, or laparoscopic appendectomy, laparoscopic colon resection or laparoscopic partial gastrectomy.
Product code
LFL Instrument, Ultrasonic Surgical
Reason for recall
A faulty footswitch may result in unintended ultrasonic fragmentation activation. There are two scenarios when the failure mode may occur: 1) Ultrasonic fragmentation may be inadvertently activated at start-up or at the initial use once the footswitch is connected to the console and after the user primes the system, performs a handpiece test, and enters the Main Screen. 2) Ultrasonic fragmentation may remain activated during surgical use once the foot pedal is disengaged slowly.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Integra I sent an Urgent Voluntary Medical Device Removal Field Safety Notice letter dated September 26, 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to check the footswitch you received with your purchase of the CUSA Clarity Console, Catalog # C7000for the lot number on the back of the footswitch. Reference pictures are provided on the next page. Compare it to the affected footswitch lot numbers on the next page. If you do have an affected lot number, replace this footswitch with the one that has been provided to you by your sales representative. If you have not yet been provided with the replacement footswitch, please contact your sales representative as soon as possible for a replacement footswitch. 2.Complete the information on the form and either provide it to your sales representative or email or fax as indicated on the form. Keep a copy of the form for your records. For questions regarding the instructions contact SSS E&L at [email protected] or 888-419-0625 option 2.
Quantity in commerce
155
Distribution
Worldwide Distribution - US Nationwide in 33 states: AL, AR, CA, CO, CT, FL, GA, IA, ID, IL, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NM, NY, OH, OK, OR, PA, TN, TX, VA, WA, WI in the countries of Greece, South Africa, France, Germany, Italy, United Kingdom, Sweden, Netherlands, and Dominican Republic.
Product codes and lots
Lot #1885744; Lot #1923802; Lot #1985221; Lot #2057996; Lot #2109587; Lot #2118507

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0694-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0694-2018Class IITerminatedVoluntary: Firm initiated