Z-0951-2018Class II
Bentson Cerebral Wire Guide
Cook Incorporated / initiated 2017-10-13 / Terminated / root cause: labeling and UDI
Cook Inc. began a recall of Bentson Cerebral Wire Guide on . FDA classified it as Class II and gives the reason as "Label does not state that the product is heparin-coated". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0951-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cook Incorporated
- Product
- Bentson Cerebral Wire Guide
- Product code
- DQX Wire, Guide, Catheter
- Reason for recall
- Label does not state that the product is heparin-coated.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- On October 31, 2017 an URGENT MEDICAL DEVICE CORRECTION was issued to customers requesting the following, including a corrective action: Step 1: Determine if your product is included in the affected lot ranges by comparing your lot number to the listing provided. " If your product is not affected, complete the Acknowledgement and Receipt Form. " If your product is affected, proceed to Step 2. Step 2: Remove a Heparin Coated label from the label sheet provided and apply to the top right corner of your affected unit. o If you need additional labels, please contact Stericycle at 855.215.4967. If you would like assistance, representatives are available to support you. To request assistance, please contact Stericycle at 855.215.4967
- Quantity in commerce
- 45
- Distribution
- Nationally
- Product codes and lots
- Catalog # THSFB-35-145-BH Affected lots to include: F3813488 - F4859795 NS5327890 - NS6060291
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0947-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0948-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0949-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0950-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0951-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0952-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0953-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0954-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0955-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0956-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0957-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0958-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0959-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0960-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0961-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0962-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0963-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0964-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0965-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0966-2018 | Class II | Terminated | Voluntary: Firm initiated |