510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0647-2018Class II

Pressure Injectable ARROWg+ard Blue PLUS Multi Lumen Central Venous Catheterization Kit

Arrow International LLC (subsidiary of Teleflex Incorporated) / initiated 2017-10-17 / Terminated / root cause: manufacturing process

Arrow International Inc began a recall of Pressure Injectable ARROWg+ard Blue PLUS Multi Lumen Central Venous Catheterization Kit; Product Code: ASK 42703 PUHC; Exp. Dates Feb 2018 - Nov 2018 on . FDA classified it as Class II and gives the reason as "These finished good kits may contain the incorrect Springwire Guide (SWG)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0647-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Arrow International LLC (subsidiary of Teleflex Incorporated)
Product
Pressure Injectable ARROWg+ard Blue PLUS Multi Lumen Central Venous Catheterization Kit; Product Code: ASK 42703 PUHC; Exp. Dates Feb 2018 - Nov 2018
Product code
FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Reason for recall
These finished good kits may contain the incorrect Springwire Guide (SWG).
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On October 17, 2017 an Urgent Medical Device recall letter from Arrow International was issued to customers requesting that customers review their inventory for product, which should immediately be discontinued and quarantined. The products are then to be returned to the firm. Questions or concerns can be directed to: Customer Service at 18662466990.
Quantity in commerce
5,926 total
Distribution
Distribution US Nationwide
Product codes and lots
Lot number: 23F16M0054

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0647-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0647-2018Class IITerminatedVoluntary: Firm initiated
Z-0648-2018Class IITerminatedVoluntary: Firm initiated
Z-0649-2018Class IITerminatedVoluntary: Firm initiated
Z-0650-2018Class IITerminatedVoluntary: Firm initiated
Z-0651-2018Class IITerminatedVoluntary: Firm initiated
Z-0652-2018Class IITerminatedVoluntary: Firm initiated
Z-0653-2018Class IITerminatedVoluntary: Firm initiated
Z-0654-2018Class IITerminatedVoluntary: Firm initiated
Z-0655-2018Class IITerminatedVoluntary: Firm initiated
Z-0656-2018Class IITerminatedVoluntary: Firm initiated
Z-0657-2018Class IITerminatedVoluntary: Firm initiated
Z-0658-2018Class IITerminatedVoluntary: Firm initiated