Silicone Fluted Round Drain 24Fr (hubless) & Connector, Sterile. Catalog Number JP-2234. Used for the...
Degania Silicone, Ltd. began a recall of Silicone Fluted Round Drain 24Fr (hubless) & Connector, Sterile. Catalog Number JP-2234. Used for the drainage of body fluids for surgical use on . FDA classified it as Class III and gives the reason as "Sterile pouches are missing labels on them. The boxes themselves are properly labeled. The issue is related to the production lot P1607442 only". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0899-2018
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Degania Silicone, Inc.
- Product
- Silicone Fluted Round Drain 24Fr (hubless) & Connector, Sterile. Catalog Number JP-2234. Used for the drainage of body fluids for surgical use.
- Product code
- GBX Catheter, Irrigation
- Reason for recall
- Sterile pouches are missing labels on them. The boxes themselves are properly labeled. The issue is related to the production lot P1607442 only.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- An Urgent Medical Device Correction letter, dated October 03, 2017, was sent to the customer. The letter identified the affected device and the reason for the recall. The letter instructed the consignee to check its stock in order to remove all mislabeled product from the market and eliminate potential risk of misuse. The customer was instructed to return the completed Confirmation Sheet to Degania Silicone, as well as any unlabeled product. Questions regarding the recall should be directed to Katia Kayam, Customer Complaints Officer at [email protected].
- Quantity in commerce
- 4,000 units
- Distribution
- US Distribution to Puerto Rico.
- Product codes and lots
- Lot No P1607442.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0899-2018 | Class III | Terminated | Voluntary: Firm initiated |