510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1010-2018Class II

NaturaLyte Liquid Acid Concentrate 2.0 mEq/L, Model # 08-2251-0 3.43 liter bottle This concentrate is...

Fresenius Medical Care Renal Therapies Group, LLC / initiated 2017-09-29 / Terminated / root cause: labeling and UDI

Fresenius Medical Care Renal Therapies Group, LLC began a recall of NaturaLyte Liquid Acid Concentrate 2.0 mEq/L, Model # 08-2251-0 3.43 liter bottle This concentrate is formulated to be used in conjunction with Fresenius Renal... on . FDA classified it as Class II and gives the reason as "The label specifies the potassium concentration is 2.0 mEq/L. Some of the product in the affected lot was found to have a potassium concentration of 2.25 mEq/L". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1010-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Product
NaturaLyte Liquid Acid Concentrate 2.0 mEq/L, Model # 08-2251-0 3.43 liter bottle This concentrate is formulated to be used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
Product code
KPO Dialysate Concentrate For Hemodialysis (Liquid Or Powder)
Reason for recall
The label specifies the potassium concentration is 2.0 mEq/L. Some of the product in the affected lot was found to have a potassium concentration of 2.25 mEq/L.
Root cause tag
labeling and UDI
FDA root cause
Release of Material/Component prior to receiving test results
Action
The firm, Fresenius Medical Care, sent an "URGENT - MEDICAL DEVICE RECALL" letter dated 10/4/2017 to its Consignees. The letter described the product, problem and actions to be taken. The Consignees are instructed to: " Immediately examine your stock to determine whether you have any Naturalyte Liquid Acid Concentrate from the lot listed. " If any product of this lot is found, discontinue use immediately. " Place all units in a secure, segregated area. " Contact your FMCRTG, LLC Customer Service Team at 1-800-323-5188 for instructions on how to return the recalled product. " Promptly fill out and return the attached Reply Form via fax to: 781-699-9796 ATTN: Quality Department or SCAN and email to: [email protected].' and "Route the letter to others in your organization that may have any of the affected product that is subject to this recall. Please note that the Naturalyte Liquid Acid Concentrate matching the Lot Number listed in the notification is the only lot subject to this recall. If you have any additional medical concerns or questions, please contact Medical Information and Communication: 855-616-2309 or Website: www.fresenius-medinfo.com.
Quantity in commerce
598 cases
Distribution
US Distribution to states of: CA, FL, KS, LA, and TX.
Product codes and lots
Lot # 17KTAC076; UDI-00840861101696

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1010-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1010-2018Class IITerminatedVoluntary: Firm initiated