Z-0829-2018Class II

2008T, Hemodialysis Delivery System Product Indicated for acute and chronic dialysis therapy

Fresenius Medical Care Renal Therapies Group, LLC / initiated 2017-10-05 / Terminated / root cause: software

Fresenius Medical Care Renal Therapies Group, LLC began a recall of 2008T, Hemodialysis Delivery System Product Indicated for acute and chronic dialysis therapy on . FDA classified it as Class II and gives the reason as "While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0829-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Product
2008T, Hemodialysis Delivery System Product Indicated for acute and chronic dialysis therapy.
Product code
KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Reason for recall
While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software anomaly is also applicable to the 2008T Machines that contained the 2.63 Bug Fix per DCAF 17-088. 2008T Upgrade kits that contained the 2.63 Bug Fix are also affected. Additionally, SW version 2.64 (in design freeze) is also impacted by this anomaly.
Root cause tag
software
FDA root cause
Software Design Change
Action
Fresenius Medical Care sent an Urgent Medical Device Correction letter dated April 24, 2017 to affected customers. the letter identified the affected product, problem and actions to be taken. For questions contact Technical Services at 1-800-227-2572.
Quantity in commerce
15 machines
Distribution
US Nationwide Distribution
Product codes and lots
2008T machines: 190766 2008T Hemodialysis System w/Bibag 190895 2008T GEN 2 Bibag without CDX  190766 = 00840861100910 (+Serial Number) 190895 = 00840861100927 (+Serial Number)ux=&Su

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0829-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0829-2018Class IITerminatedVoluntary: Firm initiated