510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1071-2018Class II

TunneLoc Tibial Fixation Device

Zimmer GmbH / initiated 2017-10-11 / Terminated / root cause: supplier

Zimmer Biomet, Inc. began a recall of TunneLoc Tibial Fixation Device Product Usage: To provide fixation of soft tissue grafts within the tibial tunnel during anterior cruciate ligament (ACL)... on . FDA classified it as Class II and gives the reason as "Gamma sterilization validation testing indicates the sterilization dose for the product may not be adequate". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.

Recall record

Recall number
Z-1071-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
TunneLoc Tibial Fixation Device Product Usage: To provide fixation of soft tissue grafts within the tibial tunnel during anterior cruciate ligament (ACL) and/or posterior cruciate ligament (PCL) reconstruction.
Product code
MBI Fastener, Fixation, Nondegradable, Soft Tissue
Reason for recall
Gamma sterilization validation testing indicates the sterilization dose for the product may not be adequate.
Root cause tag
supplier
FDA root cause
Vendor change control
Action
Zimmer Biomet sent an URGENT MEDICAL DEVICE RECALL letter dated October 16, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to review inventory and quarantine affected product, which will be returned to Biomet through a sales representative. Questions or concerns can be directed to 574-371-3071
Quantity in commerce
14,461
Distribution
US Nationwide Distribution
Product codes and lots
906512, 906513, 906514, 906515

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1071-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1071-2018Class IITerminatedVoluntary: Firm initiated
Z-1072-2018Class IITerminatedVoluntary: Firm initiated