TunneLoc Tibial Fixation Device (Within an ACL Kit)
Zimmer Biomet, Inc. began a recall of TunneLoc Tibial Fixation Device (Within an ACL Kit) Product Usage: To provide fixation of soft tissue grafts within the tibial tunnel during anterior cruciate... on . FDA classified it as Class II and gives the reason as "Gamma sterilization validation testing indicates the sterilization dose for the product may not be adequate". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.
Recall record
- Recall number
- Z-1072-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- TunneLoc Tibial Fixation Device (Within an ACL Kit) Product Usage: To provide fixation of soft tissue grafts within the tibial tunnel during anterior cruciate ligament (ACL) and/or posterior cruciate ligament (PCL) reconstruction.
- Product code
- MBI Fastener, Fixation, Nondegradable, Soft Tissue
- Reason for recall
- Gamma sterilization validation testing indicates the sterilization dose for the product may not be adequate.
- Root cause tag
- supplier
- FDA root cause
- Vendor change control
- Action
- Zimmer Biomet sent an URGENT MEDICAL DEVICE RECALL letter dated October 16, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to review inventory and quarantine affected product, which will be returned to Biomet through a sales representative. Questions or concerns can be directed to 574-371-3071
- Distribution
- US Nationwide Distribution
- Product codes and lots
- The kits that include the TunneLoc products, ACL in a box, PNs 906532, 906533, 906534, and 906535 are packed in a box alongside the Instructions for use and is distributed non-sterile .
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1071-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1072-2018 | Class II | Terminated | Voluntary: Firm initiated |