Z-1149-2018Class II

Wash Kit (REF 10680) Lots 16279, 17045 and 17215 Medica corporation is marketing a wash solution Chemistry...

Medica Corp. / initiated 2017-11-21 / Terminated

Medica Corporation began a recall of Wash Kit (REF 10680) Lots 16279, 17045 and 17215 Medica corporation is marketing a wash solution Chemistry Wash1 , REF # 10680 for use with Medica s EasyRA... on . FDA classified it as Class II and gives the reason as "The firm has initiated a field recall of the Wash1 Kit REF 10680 due to reported inconsistent Lipase values when analyzing patient samples. The firm's internal investigation identified the underlying cause to be an...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1149-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medica Corp.
Product
Wash Kit (REF 10680) Lots 16279, 17045 and 17215 Medica corporation is marketing a wash solution Chemistry Wash1 , REF # 10680 for use with Medica s EasyRA clinical chemistry analyzer. The solution is used prior to a lipase (LIP) assay if an interferant assay (TRIG)is part of the work list. The major interferant for Lipase is the Triglycerides (TRIG) assay since one of the reagents contains lipase as an active ingredient.
Product code
CHI Lipase-Esterase, Enzymatic, Photometric, Lipase
Reason for recall
The firm has initiated a field recall of the Wash1 Kit REF 10680 due to reported inconsistent Lipase values when analyzing patient samples. The firm's internal investigation identified the underlying cause to be an ineffective wash solution resulting in an interference carryover from the triglyceride (TRIG) assay if they are on the same worklist.
FDA root cause
Under Investigation by firm
Action
Firm emailed all consignees and instructed them to remove/discard affected wedges (bottles) and contact Medica for immediate replacement 1 Immediately discontinue use of the Wash1 wedges (REF 10680) (Lot# listed above) and discard any remaining kits/wedges. 2. Please discuss possible concerns about lipase results achieved using these lots of Wash wedges with your laboratory manager and/or clinician in order to decide on appropriate measures. Report any concerns to Medica for evaluation and review. 3.Submit claims for replacement of affected product to your authorized Medica EasyRA dealer. For further questions, please call (781) 541-7560.
Quantity in commerce
392
Distribution
Worldwide and US Distribution.
Product codes and lots
Lots 16279, 17045 and 17215