510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1009-2018Class II

fastener, fixation, nondegradable, soft tissue

Zimmer GmbH / initiated 2017-10-25 / Terminated / root cause: manufacturing process

Zimmer Biomet, Inc. began a recall of fastener, fixation, nondegradable, soft tissue on . FDA classified it as Class II and gives the reason as "It was identified that the product was not assembled correctly causing the implant shaft to shift and fall out during the opening of the device, which potentially could break the device". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1009-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
fastener, fixation, nondegradable, soft tissue
Product code
MBI Fastener, Fixation, Nondegradable, Soft Tissue
Reason for recall
It was identified that the product was not assembled correctly causing the implant shaft to shift and fall out during the opening of the device, which potentially could break the device.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
In October of 2017 an URGENT MEDICAL DEVICE RECALL LETTER was issued to customers requesting that any remaining product be quarantined and returned to recalling firm. Questions or concerns can be directed to 574-371-3071.
Quantity in commerce
429 units
Distribution
TX, NC, FL, PA, IN, WI, VA, NY, NJ, AL, CT
Product codes and lots
Item # 110019010 Lot #'s 017350 141990 142000 142020 167190 240720 240740 240770 257970 398710 398730 398740 398750 421780 421790 421800 465910 504120 813190 940370

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1009-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1009-2018Class IITerminatedVoluntary: Firm initiated