LeadCare II Blood Lead Analyzer Part Numbers: 70-6760, 70-6529-R, 70-6760-P, 70-6760-R. Intended for in vitro...
Magellan Diagnostics, Inc. began a recall of LeadCare II Blood Lead Analyzer Part Numbers: 70-6760, 70-6529-R, 70-6760-P, 70-6760-R. Intended for in vitro (external) use only on . FDA classified it as Class II and gives the reason as "The incorrect placement of AA batteries into certain LeadCare II Blood Lead Analyzer may cause the batteries to overheat, and the chance that a user could burn their fingers if the batteries are removed while hot". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1295-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Magellan Diagnostics, Inc.
- Product
- LeadCare II Blood Lead Analyzer Part Numbers: 70-6760, 70-6529-R, 70-6760-P, 70-6760-R. Intended for in vitro (external) use only.
- Product code
- DOF Lead, Atomic Absorption
- Reason for recall
- The incorrect placement of AA batteries into certain LeadCare II Blood Lead Analyzer may cause the batteries to overheat, and the chance that a user could burn their fingers if the batteries are removed while hot.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- An Urgent Medical Device Correction letter, dated January 9, 2018, was sent to customers. The letter identified the affected device and provided a description of the issue. If two of the four AA batteries are inserted incorrectly, the batteries may overheat, causing a minor burn if the user opens the battery compartment and tries to remove the hot batteries. Customers are to follow the steps outlined in the letter, including placing the enclosed label on the upper back panel of the analyzer to remind users who are changing batteries of the importance of correct placement. Proper battery placement should be provided in annual training. Customers are to complete and return the acknowledgement card. Please direct any questions to 800-275-0102 or [email protected].
- Quantity in commerce
- 6,143
- Distribution
- Worldwide Distribution
- Product codes and lots
- Part Numbers: 70-6760, 70-6529-R, 70-6760-P, and 70-6760-R. All LeadCare II and LeadCare Plus Blood Lead Analyzers with battery boards versions B and O.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1295-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1296-2018 | Class II | Terminated | Voluntary: Firm initiated |