Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1068-2018Class II

STA - Neoplastine Cl (ref. 00605)

Diagnostica Stago, Inc. / initiated 2017-10-25 / Terminated / root cause: design

Diagnostica Stago, Inc. began a recall of STA - Neoplastine Cl (ref. 00605) Product Usage: Manual or automated determination of the prothrombin time (PT) on . FDA classified it as Class II and gives the reason as "Confirmed defect of homogeneity where some reagent vials will give prolonged Prothrombin Time (decreased PT%)". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1068-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Diagnostica Stago, Inc.
Product
STA - Neoplastine Cl (ref. 00605) Product Usage: Manual or automated determination of the prothrombin time (PT).
Product code
GJS Test, Time, Prothrombin
Reason for recall
Confirmed defect of homogeneity where some reagent vials will give prolonged Prothrombin Time (decreased PT%).
Root cause tag
design
FDA root cause
Device Design
Action
Stago sent an Urgent Field Safety Notice letter dated October 25, 2017 to their customers. The letter identified the affected product, problem and actions to be taken: Customers were instructed to: If it is not already done, run a Quality Control test at every change of vial (if you have affected lots of product). Return to company, by fax or email, an acknowledgement form confirming that the recall letter was read by the receiving company. For questions contact the Stago Hotline (1-800-725-0607).
Distribution
Nationally
Product codes and lots
Lot 251064 (Exp: 09-2018) Lot 251725 (Exp: 02-2019)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1068-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1066-2018Class IITerminatedVoluntary: Firm initiated
Z-1067-2018Class IITerminatedVoluntary: Firm initiated
Z-1068-2018Class IITerminatedVoluntary: Firm initiated
Z-1069-2018Class IITerminatedVoluntary: Firm initiated