Z-0712-2018Class II

ProLite Mesh - mesh, surgical, polymeric ProLite Mesh is intended for use in soft tissue deficiencies...

Atrium Medical Corporation / initiated 2017-11-22 / Terminated / root cause: labeling and UDI

Atrium Medical Corporation began a recall of ProLite Mesh - mesh, surgical, polymeric ProLite Mesh is intended for use in soft tissue deficiencies including hernia repair, traumatic or surgical wounds and... on . FDA classified it as Class II and gives the reason as "An incorrect ProLite Mesh configuration was packaged as product code 1010306-06, with lot number 412298. ProLite Mesh manufactured as product code 1010306-04 was inadvertently placed inside the pouch of packages labeled...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0712-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Atrium Medical Corporation
Product
ProLite Mesh - mesh, surgical, polymeric ProLite Mesh is intended for use in soft tissue deficiencies including hernia repair, traumatic or surgical wounds and chest wall reconstruction procedures requiring reinforcement with a non-absorbable supportive material.
Product code
FTL Mesh, Surgical, Polymeric
Reason for recall
An incorrect ProLite Mesh configuration was packaged as product code 1010306-06, with lot number 412298. ProLite Mesh manufactured as product code 1010306-04 was inadvertently placed inside the pouch of packages labeled with product code 1010306-06 and lot number 412298.
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
Please examine inventory to determine if you have any of the affected products listed on page 1. If so, please remove affected products, quarantine them, and place in a secure location. If you have affected product, please contact Getinge Customer Service at [email protected] for a Return Authorization (RA) and shipping instructions to return any affected product. Pack the product to be returned with the appropriate return documents, and using the shipping instructions provided, arrange for pick up with the designated delivery service provider. Please complete and sign the enclosed Urgent Medical Device Field Safety Notice to acknowledge receipt. Return the completed form to the following e-mail address: [email protected] For further questions, please call (973) 709-7654.
Quantity in commerce
60 units
Distribution
Internationally to Spain only.
Product codes and lots
Product Code/REF 1010306-06 Lot Number/LOT: 412298

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0712-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0712-2018Class IITerminatedVoluntary: Firm initiated