Z-1130-2018Class II

Situate Gauze RF and X-Ray detectable. Product Number: G0804-16P02CN-1 Situate premium cotton are single use...

Medtronic, Inc. / initiated 2017-10-12 / Terminated / root cause: manufacturing process

Covidien LLC began a recall of Situate Gauze RF and X-Ray detectable. Product Number: G0804-16P02CN-1 Situate premium cotton are single use cotton disposable that must be used with the... on . FDA classified it as Class II and gives the reason as "This voluntary recall is being conducted due to the potential for failure to detect the affected Situate" gauze - RF and Xray detectable products by the Situate" detection console due to a manufacturing error". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1130-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Situate Gauze RF and X-Ray detectable. Product Number: G0804-16P02CN-1 Situate premium cotton are single use cotton disposable that must be used with the Situate delivery system. The disposable have an attached RF Tag, which is an electrically passive inductor like device that is sewn into the green pouch that you will see on all Situate premium cotton. When activated by magnetic impulses from the scanning devices, the tag resonates a signal that allows for the detection of these cotton disposable.
Product code
GDY Gauze/Sponge, Internal, X-Ray Detectable
Reason for recall
This voluntary recall is being conducted due to the potential for failure to detect the affected Situate" gauze - RF and Xray detectable products by the Situate" detection console due to a manufacturing error.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
All consignees were notified via FedEx or certified mail on October 12, 2017 and the letter 'Urgent Field Corrective action Notice' informs customers of the potential for failure to detect the affected Situate" gauze RF and X-Ray detectable products by the Situate" detection console and the actions they should take. Customers are requested to acknowledge receipt and understanding of the Urgent Field Correction Action Notice. For further questions, please call (203) 492-5000.
Quantity in commerce
20,800
Distribution
U.S.
Product codes and lots
Lot Number: 170704A UDI: 20884521700861

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1130-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1130-2018Class IITerminatedVoluntary: Firm initiated