Z-1285-2018Class II
Radial Artery Catheterization Kit; Model: AK-04122
Arrow International LLC (subsidiary of Teleflex Incorporated) / initiated 2017-10-26 / Terminated / root cause: manufacturing process
Arrow International Inc began a recall of Radial Artery Catheterization Kit; Model: AK-04122 on . FDA classified it as Class II and gives the reason as "The povidone iodine prep pads included in certain kits and trays do not meet the requirements for a 36-month shelf life and should be labeled with a 24-month shelf life". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1285-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Arrow International LLC (subsidiary of Teleflex Incorporated)
- Product
- Radial Artery Catheterization Kit; Model: AK-04122
- Product code
- DQX Wire, Guide, Catheter
- Reason for recall
- The povidone iodine prep pads included in certain kits and trays do not meet the requirements for a 36-month shelf life and should be labeled with a 24-month shelf life.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Teleflex/Arrow sent a letter dated October 26th, 2017. Customers are instructed to discontinue use and quarantine any products with the affected product codes and lot numbers. There is a Recall Acknowledgment Form that is requested to be completed that will allow the return and replacement of the affected product. Consignees without affected product on hand are also asked to return the form.
- Quantity in commerce
- 570
- Distribution
- US nationwide distribution, including Puerto Rico.
- Product codes and lots
- Lots: 13F16H0329, 13F17F0074, 13F17G0237; Expiration dates: October 2018 through March 2020
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1281-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1282-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1283-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1284-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1285-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1286-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1287-2018 | Class II | Terminated | Voluntary: Firm initiated |