Z-1257-2018Class II

Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 42mm, REF 804-07-420

Encore Medical, LP / initiated 2018-01-29 / Completed / root cause: design

Encore Medical, Lp began a recall of Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 42mm, REF 804-07-420 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder... on . FDA classified it as Class II and gives the reason as "It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540) is much larger on the...". As of , the recall status on record is Completed, and MDM tags the root cause as design.

Recall record

Recall number
Z-1257-2018
Classification
Class II
Status
Completed
Initiated
Posted
Recalling firm
Encore Medical, LP
Product
Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 42mm, REF 804-07-420 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder system indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction.
Product code
KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
Reason for recall
It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540) is much larger on the trails than on the implants
Root cause tag
design
FDA root cause
Device Design
Action
DJO Global sent an Urgent Field Safety Notice letter dated January 29, 2018, to affected customers. The letter identified the affected product, problem and actions to be taken. For questions call (512) 834-6255
Quantity in commerce
92 units
Distribution
US Nationwide Distribution
Product codes and lots
All Serial Numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1257-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1256-2018Class IICompletedVoluntary: Firm initiated
Z-1257-2018Class IICompletedVoluntary: Firm initiated
Z-1258-2018Class IICompletedVoluntary: Firm initiated
Z-1259-2018Class IICompletedVoluntary: Firm initiated
Z-1260-2018Class IICompletedVoluntary: Firm initiated