Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 50mm, REF 804-07-500
Encore Medical, Lp began a recall of Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 50mm, REF 804-07-500 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder... on . FDA classified it as Class II and gives the reason as "It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540) is much larger on the...". As of , the recall status on record is Completed, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1259-2018
- Classification
- Class II
- Status
- Completed
- Initiated
- Posted
- Recalling firm
- Encore Medical, LP
- Product
- Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 50mm, REF 804-07-500 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder system indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction.
- Product code
- KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
- Reason for recall
- It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540) is much larger on the trails than on the implants
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- DJO Global sent an Urgent Field Safety Notice letter dated January 29, 2018, to affected customers. The letter identified the affected product, problem and actions to be taken. For questions call (512) 834-6255
- Quantity in commerce
- 92 units
- Distribution
- US Nationwide Distribution
- Product codes and lots
- All Serial Numbers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1256-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1257-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1258-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1259-2018 | Class II | Completed | Voluntary: Firm initiated | ||
| Z-1260-2018 | Class II | Completed | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.