Z-1078-2018Class II
Skin Staple Remover
Centurion Medical Products, LP / initiated 2017-10-23 / Terminated / root cause: manufacturing process
Centurion Medical Products Corporation began a recall of SKIN STAPLE REMOVER on . FDA classified it as Class II and gives the reason as "Convenience Kit contains a PVP Prep Pad, received by a supplier, that does not meet the iodine assay level required to support the 36 month shelf life". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1078-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Centurion Medical Products, LP
- Product
- Skin Staple Remover
- Product code
- MCZ Suture Removal Kit
- Reason for recall
- Convenience Kit contains a PVP Prep Pad, received by a supplier, that does not meet the iodine assay level required to support the 36 month shelf life.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On October 16, 2017 a PRODUCT CORRECTION letter was issued to customers notifying them that if they had any kits in their inventory, Centurion will supply labels to affix to the front of each kit instructing the user to discard the prep pad upon opening the kit. Other components of the kit may continue to be used and are NOT within the scope of this recall.
- Distribution
- Nationally
- Product codes and lots
- 44105 Lot codes: 2015083101 2016101701 2017082801 2017030601 2016072501
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1078-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1079-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1080-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1081-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1082-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1083-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1084-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1085-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1086-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1087-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1088-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1089-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1090-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1091-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1092-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1093-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1094-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1095-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1096-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1097-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1098-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
