Z-1913-2018Class II
Philips HeartStart FRx AED
Philips North America LLC / initiated 2018-03-14 / Terminated / root cause: manufacturing process
Philips Electronics North America Corp. began a recall of Philips HeartStart FRx AED on . FDA classified it as Class II and gives the reason as "An issue with one of the electric components (a resistor) could result in the device not delivering a shock when needed". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1913-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips HeartStart FRx AED.
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- An issue with one of the electric components (a resistor) could result in the device not delivering a shock when needed.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Letters dated 2/16/2018 for U.S. customers were issued via certified mail on 3/14/2018. Letters to customers outside of the U.S. were localized in conjunction with the local Philips market organizations and are in progress.
- Quantity in commerce
- Total for FRx and HS1 - 660,519 defibrillators
- Distribution
- Distribution was nationwide and to Puerto Rico. There was also government/military distribution. Foreign distribution was made to the following countries: Canada, Afghanistan, Algeria, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Botswana, Brazil, Brunei Darussalam, Cambodia, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Fiji, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Malaysia, Malta, Mauritius, Mexico, Morocco, Namibia, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Niger, Norway, Oman, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Suriname, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirate, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen, Zambia, and Zimbabwe.
- Product codes and lots
- Manufactured between September 2002 and February 2013. The year of manufacture can be identified by the 2nd and 3rd characters in the serial number on the back of the AED in the following ranges: Home/Onsite: A02x-xxxxx through A13B-xxxxx; FRx: B04x-xxxxx through B13B-xxxxx NOTE: If the device was manufactured in 2013 and the 4th digit is the letter "C" or later (D, E, F...), it is not covered by the recall (i.e. A13G-02375). In some cases, a device within the date range covered by the notification did not contain a resistor previously associated with a failure. In that case, a notification was not sent. Confirmation as to whether the device contains the resistor at issue can be done by contacting the firm.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1913-2018 | Class II | Terminated | FDA Mandated | ||
| Z-1914-2018 | Class II | Terminated | FDA Mandated |