Argyle Suction Tubing (Product # 8888301515) Argyle Suction Tubing is an extruded tube used as an extension...
Medtronic Minimally Invasive Therapies Group began a recall of Argyle Suction Tubing (Product # 8888301515) Argyle Suction Tubing is an extruded tube used as an extension for different types of devices in an operating... on . FDA classified it as Class II and gives the reason as "Incorrect packaging. In one lot of Argyle Suction Tubing, the pouch may be incorrectly labeled as the Salem Sump Dual Lumen Stomach Tube. This product may have been distributed within the case of Argyle Suction Tubing". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0936-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Argyle Suction Tubing (Product # 8888301515) Argyle Suction Tubing is an extruded tube used as an extension for different types of devices in an operating room, bonded with 2 female sure grip connectors at each end of the tube to facilitate the plug/ insertion of the device and also includes a male connector that can be used for devices designed with female connectors, This product is sterile, non-conductive and latex free.
- Product code
- GAZ Tubing, Noninvasive
- Reason for recall
- Incorrect packaging. In one lot of Argyle Suction Tubing, the pouch may be incorrectly labeled as the Salem Sump Dual Lumen Stomach Tube. This product may have been distributed within the case of Argyle Suction Tubing.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- An URGENT MEDICAL DEVICE RECALL letter was issued to customers on October 24, 2017. This letter notified customers of the products subject to the recall and stated the following: The primary package labeling (Label closest to the product) may be incorrect. Customers can identify the affected product by consulting Attachment B. Please quarantine and return any unused products of the item codes and lot detailed above. Unused products from the affected item code and lots should be returned as described in the Required Actions section below. If you have distributed the Argyle" Suction tubing listed above, please promptly forward the information from this letter to those recipients. All unused products from the affected item codes and lots must be returned. If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
- Quantity in commerce
- 58,954
- Distribution
- USA (nationwide) Distribution
- Product codes and lots
- 710016864X
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0936-2018 | Class II | Terminated | Voluntary: Firm initiated |