AltiVate Reverse INSTRUMENTATION, HUMERAL SOCKET SHELL TRIAL, STANDARD, REF 804-06-152, Qty: 01, Encore...
Encore Medical, Lp began a recall of AltiVate Reverse INSTRUMENTATION, HUMERAL SOCKET SHELL TRIAL, STANDARD, REF 804-06-152, Qty: 01, Encore Medical, LP Orthopedic surgical instrument. Product... on . FDA classified it as Class II and gives the reason as "Broken or missing screws and retaining clips. This may be due to screw tolerance and material strength of the trial". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1241-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Encore Medical, LP
- Product
- AltiVate Reverse INSTRUMENTATION, HUMERAL SOCKET SHELL TRIAL, STANDARD, REF 804-06-152, Qty: 01, Encore Medical, LP Orthopedic surgical instrument. Product Usage: The humeral socket shell trial is used to replicate/trial the shell geometry of the AltiVate Reverse Standard Shell Humeral Stem implants. It assembles to the humeral broaches by a screw in order to replicate/trial the entire humeral stem implant geometry prior to implantation.
- Product code
- PHX Shoulder Prosthesis, Reverse Configuration
- Reason for recall
- Broken or missing screws and retaining clips. This may be due to screw tolerance and material strength of the trial.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm disseminated their field safety notices on 01/30/2018 by email. The letter identified the affected product, problem and actions to be taken. The notices requested the trails be returned for replacement.
- Quantity in commerce
- 387 units
- Distribution
- US Nationwide Distribution
- Product codes and lots
- Lot Codes: 183312L03, 200006L01, 224643L01
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1241-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.