Z-0976-2018Class II

IMRIS T2X Table, Part Number 109682-600

IMRIS Imaging, Inc. / initiated 2018-02-20 / Terminated / root cause: design

Deerfield Imaging, Inc. began a recall of IMRIS T2X Table, Part Number 109682-600 on . FDA classified it as Class II and gives the reason as "The rotational locking mechanisms on the ORT 100/ T2X tables can become disengaged or unlocked while in use which could potentially result in unintended movement of the table". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0976-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
IMRIS Imaging, Inc.
Product
IMRIS T2X Table, Part Number 109682-600
Product code
FQO Table, Operating-Room, Ac-Powered
Reason for recall
The rotational locking mechanisms on the ORT 100/ T2X tables can become disengaged or unlocked while in use which could potentially result in unintended movement of the table.
Root cause tag
design
FDA root cause
Device Design
Action
The firm disseminated the notices on 02/23/2018 and followed with emails on 02/26/2018. The notices stated: "To minimize the risk of injury, IMRIS is taking the following corrective actions: " ORT100 Table: The rotational locking mechanism shall be replaced in all ORT100 tables that do not contain the most current design. " T2X Table: Tables will have their rotational lock inspected. If components are worn or performance is other than intended, the rotational lock assembly will be serviced or replaced entirely. IMRIS has identified that your facility has one or more of these OR Tables which require evaluation. Please confirm details on page 2 and return your response to IMRIS.IMRIS Customer Service will contact you to implement this corrective action. IMRIS recommends these preventive actions: " Per the ORT Operator Manuals the table should be tested daily in all standard positions and inspected for anomalies including the fit of the covers, back section, headrest, pendant connection and swivel release pedal free engagement " During usage, verify the table is locked by manually pushing the table to make sure it does not move rotationally. " If the rotational lock is observed to be not functioning properly, please contact IMRIS Customer Service. The appropriate competent authorities have been notified of this corrective action. IMRIS is committed to patient and user safety, and apologizes for any inconvenience that this may cause. Please contact customer support via our service line at 1-866-475-0525 or Email: customersupport@i mris.com if you have any questions or concerns."
Quantity in commerce
16 tables
Distribution
US, Canada, Australia, China, Qatar
Product codes and lots
Serial Numbers: 10000239, 10001306, 10000983, 10000982, 10000233, 10001119, 10001192, 10002375, 10003623, 10003623, 05 002, 10003634, 10001118, 10001188, 10001340, 10003638

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0976-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0975-2018Class IITerminatedVoluntary: Firm initiated
Z-0976-2018Class IITerminatedVoluntary: Firm initiated