Z-1294-2018Class II
Alpha Air 6 ENT Microscope Microscope examination of the human ear, nose, and throat
Seiler Instrument & Mfg. Co.,Inc. / initiated 2017-01-01 / Open, Classified
Seiler Instr. & Mfg. Co. Inc. began a recall of Alpha Air 6 ENT Microscope Microscope examination of the human ear, nose, and throat on . FDA classified it as Class II and gives the reason as "The microscope may tip over if attachments are fully extended in one line and there is an attempt to move the device not by the handles mounted on the post". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1294-2018
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Seiler Instrument & Mfg. Co.,Inc.
- Product
- Alpha Air 6 ENT Microscope Microscope examination of the human ear, nose, and throat.
- Product code
- EPT Microscope, Surgical
- Reason for recall
- The microscope may tip over if attachments are fully extended in one line and there is an attempt to move the device not by the handles mounted on the post.
- FDA root cause
- Use error
- Action
- The customer was notified via email on about January 4, 2017. Instructions included that the scope only tips if the user has it extended all the way out and is pulling on the handles of the scope. The end user must not pull on the handles of the scope when it is extended. They were also instructed to educate personnel on this matter. For further questions, please call (314) 968-2282.
- Quantity in commerce
- 43 units, but 1 affected
- Distribution
- Worldwide Distribution - US Distribution to the states of : CA, CO, FL, IL, MN, NV, OK, TX., and to the countries of : Canada, China, Ecuador, England, Greece, Guatemala, Jordan, Lebanon, Mexico, Romania, Russia, South Korea, Taiwan, Japan and United Arab Emirates.
- Product codes and lots
- Model Number 6 ENT
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1294-2018 | Class II | Ongoing | Voluntary: Firm initiated |