Z-1087-2018Class II
Minor Dressing Change Tray
Centurion Medical Products, LP / initiated 2017-10-23 / Terminated / root cause: manufacturing process
Centurion Medical Products Corporation began a recall of MINOR DRESSING CHANGE TRAY on . FDA classified it as Class II and gives the reason as "Convenience Kit contains a PVP Prep Pad, received by a supplier, that does not meet the iodine assay level required to support the 36 month shelf life". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1087-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Centurion Medical Products, LP
- Product
- Minor Dressing Change Tray
- Product code
- LRP Tray, Surgical
- Reason for recall
- Convenience Kit contains a PVP Prep Pad, received by a supplier, that does not meet the iodine assay level required to support the 36 month shelf life.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On October 16, 2017 a PRODUCT CORRECTION letter was issued to customers notifying them that if they had any kits in their inventory, Centurion will supply labels to affix to the front of each kit instructing the user to discard the prep pad upon opening the kit. Other components of the kit may continue to be used and are NOT within the scope of this recall.
- Distribution
- Nationally
- Product codes and lots
- DT875 Lot codes: 2015102090 2015120290 2016012190 2016032290 2016052690 2016072890 2016090590 2017010490 2017020790 2017032190 2017042790
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1078-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1079-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1080-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1081-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1082-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1083-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1084-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1085-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1086-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1087-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1088-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1089-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1090-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1091-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1092-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1093-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1094-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1095-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1096-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1097-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1098-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
