Z-1341-2018Class II

AQUIOS CL Flow Cytometer System, Catalog #B30166. The AQUIOS CL Flow Cytometer system is an automated...

Siemens Healthcare Diagnostics Inc. / initiated 2018-01-29 / Terminated / root cause: software

Beckman Coulter Inc. began a recall of AQUIOS CL Flow Cytometer System, Catalog #B30166. The AQUIOS CL Flow Cytometer system is an automated analyzer that use a no-wash sample preparation... on . FDA classified it as Class II and gives the reason as "The device may process the same sample with two different sample IDs and sample information while using the single tube loader, which has the potential for erroneous results due to the mis-identification". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1341-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
AQUIOS CL Flow Cytometer System, Catalog #B30166. The AQUIOS CL Flow Cytometer system is an automated analyzer that use a no-wash sample preparation process.The firm name on the label is Beckman Coulter Ireland, Inc., Co. Clare, Ireland.
Product code
OYE Flow Cytometric Reagents And Accessories.
Reason for recall
The device may process the same sample with two different sample IDs and sample information while using the single tube loader, which has the potential for erroneous results due to the mis-identification.
Root cause tag
software
FDA root cause
Software design
Action
The recalling firm issued letters dated 1/25/2018 via email on 1/31/2018 or to some customers via regular mail during the week of 1/29/2018.
Quantity in commerce
288 devices
Distribution
Distribution was made nationwide in the U.S. and in Puerto Rico. There was also government distribution, but no military distribution. Foreign distribution was made to Angola, Australia, Belgium, Botswana, Brazil, Canada, Chile, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ghana, Hong Kong, India, Israel, Italy, Japan, Kazakhstan, Kenya, Lesotho, Malawi, Malaysia, Mexico, Namibia, Netherlands, New Zealand, Norway, Pakistan, Panama, Poland, Portugal, Romania, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, Uganda, Ukraine, United Kingdom, and Zambia.
Product codes and lots
All software versions.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1341-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1341-2018Class IITerminatedVoluntary: Firm initiated