510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1064-2018Class II

6.5mm x 125mm x 16mm Cannulated Screw, Item Number: 110009948 (UDI: (01)00887868111327 (17)270421 (10)120910)

Zimmer GmbH / initiated 2017-12-07 / Terminated / root cause: labeling and UDI

Zimmer Biomet, Inc. began a recall of 6.5mm x 125mm x 16mm Cannulated Screw, Item Number: 110009948 (UDI: (01)00887868111327 (17)270421 (10)120910) Product Usage: The 6.5/8.0mm Cannulated Screw... on . FDA classified it as Class II and gives the reason as "One lot of 3.5mm x 24mm Cortical Locking screws is labeled as 6.5mm x 125mm x 16mm Cannulated screws". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1064-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
6.5mm x 125mm x 16mm Cannulated Screw, Item Number: 110009948 (UDI: (01)00887868111327 (17)270421 (10)120910) Product Usage: The 6.5/8.0mm Cannulated Screw System is part of a series of cannulated screws ranging from 4.0 to 8.0mm. Used in Slipped capital femoral epiphysis, Pediatric femoral neck fractures, Tibial plateau fractures, SI joint disruptions, Intercondylar femur fractures, Subtalar arthrodesis, and Fixation of pelvis and iliosacral joint.
Product code
HWC Screw, Fixation, Bone
Reason for recall
One lot of 3.5mm x 24mm Cortical Locking screws is labeled as 6.5mm x 125mm x 16mm Cannulated screws.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Zimmer Biomet sent an Urgent Medical Device Recall letter dated December 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to: 1. Review this notification n and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form and send to [email protected] within three (3) days. b. Request an International Return Authorization Number by emailing [email protected] c. For each return, send a copy of Attachment 1 to [email protected]. d. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. e. Mark "RECALL" on the outside of the returned cartons. 4. Return a copy of the your Inventory Return Certification and product return form. 5. For questions call customer service at 574-374-3071.
Quantity in commerce
6
Distribution
Canada.
Product codes and lots
Lot: 120910

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1064-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1064-2018Class IITerminatedVoluntary: Firm initiated