6.5mm x 125mm x 16mm Cannulated Screw, Item Number: 110009948 (UDI: (01)00887868111327 (17)270421 (10)120910)
Zimmer Biomet, Inc. began a recall of 6.5mm x 125mm x 16mm Cannulated Screw, Item Number: 110009948 (UDI: (01)00887868111327 (17)270421 (10)120910) Product Usage: The 6.5/8.0mm Cannulated Screw... on . FDA classified it as Class II and gives the reason as "One lot of 3.5mm x 24mm Cortical Locking screws is labeled as 6.5mm x 125mm x 16mm Cannulated screws". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1064-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- 6.5mm x 125mm x 16mm Cannulated Screw, Item Number: 110009948 (UDI: (01)00887868111327 (17)270421 (10)120910) Product Usage: The 6.5/8.0mm Cannulated Screw System is part of a series of cannulated screws ranging from 4.0 to 8.0mm. Used in Slipped capital femoral epiphysis, Pediatric femoral neck fractures, Tibial plateau fractures, SI joint disruptions, Intercondylar femur fractures, Subtalar arthrodesis, and Fixation of pelvis and iliosacral joint.
- Product code
- HWC Screw, Fixation, Bone
- Reason for recall
- One lot of 3.5mm x 24mm Cortical Locking screws is labeled as 6.5mm x 125mm x 16mm Cannulated screws.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Zimmer Biomet sent an Urgent Medical Device Recall letter dated December 2017 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to: 1. Review this notification n and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form and send to [email protected] within three (3) days. b. Request an International Return Authorization Number by emailing [email protected] c. For each return, send a copy of Attachment 1 to [email protected]. d. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. e. Mark "RECALL" on the outside of the returned cartons. 4. Return a copy of the your Inventory Return Certification and product return form. 5. For questions call customer service at 574-374-3071.
- Quantity in commerce
- 6
- Distribution
- Canada.
- Product codes and lots
- Lot: 120910
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1064-2018 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.