Z-1192-2018Class II

Zimmer M/L Taper Hip Prosthesis, 7711 Series, Femoral Stem 12/14 Neck Taper Plasma Sprayed Press-Fit...

Zimmer GmbH / initiated 2018-03-01 / Terminated / root cause: labeling and UDI

Zimmer Biomet, Inc. began a recall of Zimmer M/L Taper Hip Prosthesis, 7711 Series, Femoral Stem 12/14 Neck Taper Plasma Sprayed Press-Fit Cementless Size 4 Standard Reduced Neck Length; Part... on . FDA classified it as Class II and gives the reason as "Certain packages labeled as size 7.5 Extended Offset Reduced Neck length hip implant, actually contained a Size 4 Standard Reduced Neck length taper hip implant". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1192-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer M/L Taper Hip Prosthesis, 7711 Series, Femoral Stem 12/14 Neck Taper Plasma Sprayed Press-Fit Cementless Size 4 Standard Reduced Neck Length; Part Number: 00-7711-004-10
Product code
LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Reason for recall
Certain packages labeled as size 7.5 Extended Offset Reduced Neck length hip implant, actually contained a Size 4 Standard Reduced Neck length taper hip implant.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
RISK MANAGER: 1. Review this notification and notify affected personnel of the contents. 2. Quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Certificate of Acknowledgement and send to [email protected]. DISTRIBUTOR: 1. Review this notification and notify affected team members of the contents. 2. Quarantine affected product in your inventory. 3. Return affected product from your distributorship and from affected hospitals. a. Complete Inventory Return Certification Form and send to [email protected] Mark RECALL on the outside of the returned cartons. 4. Return the Additional Accounts form to [email protected].
Quantity in commerce
24
Distribution
Distributed in the following U.S. states: AZ, CA, CT, IN, MD, MI, MN, NJ, NY, OR, WA, and Puerto Rico. Distributed in China and India.
Product codes and lots
Lot Number: 63716613, UDI: (01)00889024131606(17)270731(10)63716613; Lot Number: 63716614, UDI: (01)00889024131606(17)279731(10)63716614

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1192-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1191-2018Class IITerminatedVoluntary: Firm initiated
Z-1192-2018Class IITerminatedVoluntary: Firm initiated