ACE Trochanteric Nail System ACE TROCH NAIL 125 DEG. X 11MM ACE TROCH NAIL 130 DEG. X 11MM ACE TROCH NAIL 135...
Zimmer Biomet, Inc. began a recall of ACE Trochanteric Nail System ACE TROCH NAIL 125 DEG. X 11MM ACE TROCH NAIL 130 DEG. X 11MM ACE TROCH NAIL 135 DEG. X 11MM ACE TROCH NAIL 125 DEG. X 13MM ACE... on . FDA classified it as Class II and gives the reason as "Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1225-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- ACE Trochanteric Nail System ACE TROCH NAIL 125 DEG. X 11MM ACE TROCH NAIL 130 DEG. X 11MM ACE TROCH NAIL 135 DEG. X 11MM ACE TROCH NAIL 125 DEG. X 13MM ACE TROCH NAIL 130 DEG. X 13MM ACE TROCH NAIL 135 DEG. X 13MM The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures in osteoporotic bone of the trochanteric and diaphyseal areas, long subtrochanteric fractures, ipsilateral femoral fractures, proximal or distal nonunions and malunions and revision procedures.
- Product code
- HSB Rod, Fixation, Intramedullary And Accessories
- Reason for recall
- Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On December 22, 2017 an URGENT MEDICAL DEVICE RECALL letter was issued to customers with a complete list of products subject to recall and requesting that the customer review the notification and notify all affected personnel of the recall. If there is affected product at the facility it is requested that these units be quarantined and returned to a sales representative for return to the firm. Questions or concerns can be directed to 574-371-3071.
- Quantity in commerce
- 3454
- Distribution
- USA (nationwide)
- Product codes and lots
- 903011225 903011230 903011235 903013225 903013230 903013235
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1225-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1226-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1227-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1228-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1229-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1230-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1231-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1232-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1233-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1234-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1235-2018 | Class II | Terminated | Voluntary: Firm initiated |