Dimension Vista System LOCI 8 CALIBRATOR The LOCI 8 Calibrator is an in vitro diagnostic product for the...
Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista System LOCI 8 CALIBRATOR The LOCI 8 Calibrator is an in vitro diagnostic product for the calibration of follicle stimulating hormone (FSH)... on . FDA classified it as Class II and gives the reason as "An increase in the number of failed calibration events or negative bias with Quality Control (QC) and patient samples when using Dimension Vista LOCI 8 Calibrator lot 7CD076 to calibrate Estradiol (E2) Assay". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1184-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension Vista System LOCI 8 CALIBRATOR The LOCI 8 Calibrator is an in vitro diagnostic product for the calibration of follicle stimulating hormone (FSH), luteinizing hormone (LH), prolactin (PRL), and estradiol (E2) methods on the Dimension Vista System.
- Product code
- JIX Calibrator, Multi-Analyte Mixture
- Reason for recall
- An increase in the number of failed calibration events or negative bias with Quality Control (QC) and patient samples when using Dimension Vista LOCI 8 Calibrator lot 7CD076 to calibrate Estradiol (E2) Assay.
- Root cause tag
- manufacturing process
- FDA root cause
- Under Investigation by firm
- Action
- The firm, Siemens, issued an "Urgent Medical Device Correction (VC18-01.A.US)" letter dated 11/3/2017, to US customers on November 3, 2017 and an "Urgent Field Safety Notice (VC18-01.A.OUS)" to customers outside the US in November 2017. Customers were instructed to: " Discontinue use of and discard impacted LOCI 8 CAL lot 7CD076 " Calibrate the E2 assay using an alternate lot of LOCI 8 CAL. " Review the letter with their Medical Director. " Review their inventory of this product to determine theirs laboratory's replacement needs and to provide information to Siemens for reporting to the regulatory authorities. " Complete and return the Field Correction Effectiveness Check Form attached to the UMDR/UFSN letter within 30 days via Fax to Customer Care Center at (312) 275-7795 or send scanned copy via email to [email protected]. " Retain this letter with your laboratory records, and forward this letter to those who may have received this product. If you have any questions, please contact your Siemens Customer Care Center, your local Siemens Technical Support Representative or call recall contact at 800-441-9250; email: [email protected].
- Quantity in commerce
- 989 (681 USA/308 OUS)
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: Austria, Belgium, Canada, Denmark, France, Germany, Italy, Netherlands, Slovakia, Spain and Switzerland.
- Product codes and lots
- CATALOG #(s): KC646 Lot 7CD076- UDI# (01)00842768026492(10)7CD076(17)20180401
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1184-2018 | Class II | Terminated | Voluntary: Firm initiated |