Specimen Gate Screening Center, Part Number 5002-0500, All software versions from 1.0 to 1.8 (current...
PerkinElmer Life and Analytical Sciences, Wallac, OY began a recall of Specimen Gate Screening Center, Part Number 5002-0500, All software versions from 1.0 to 1.8 (current software version) Specimen Gate Screening Center is used... on . FDA classified it as Class II and gives the reason as "Potential errors in patient results generated by the Screening Center product that include both false negative and false positive results". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1541-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Wallac Oy
- Product
- Specimen Gate Screening Center, Part Number 5002-0500, All software versions from 1.0 to 1.8 (current software version) Specimen Gate Screening Center is used for data management of neonatal screening test results and demographics by qualified laboratory personnel in newborn screening programs.
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- Potential errors in patient results generated by the Screening Center product that include both false negative and false positive results.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- PerkinElmer sent an Urgent Medical Device Correction letter dated November 14,2017. Instructions: 1. Implement additional control measures concerning demographic information, by taking one of the following actions a. Fully enter demographics before loading specimens on instruments (i.e. specimens can be punched, but the assay plates cannot be loaded to the measurement instruments). b. If the above mentioned is not possible or if previously entered demographic information changes after loading the assay plates on the measurement instruments, recalculate all your results in Specimen Gate laboratory Result Viewer prior to assay acceptance. 2. Follow unreported specimens with available tools to avoid any delay in reporting screening results. For further questions, please call (317) 418-1735
- Quantity in commerce
- 8
- Distribution
- Worldwide Distribution - US Distribution to the state of Georgia., and to the countries of : Belgium, Canada, Denmark, Italy, and United Kingdom.
- Product codes and lots
- All software versions from 1.0 to 1.8 (current software version)