Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1242-2018Class III

Gas Module 3 monitoring spirometer, Part Number: 115-030108-00

Datascope Corp. / initiated 2017-12-01 / Terminated / root cause: design

Mindray DS USA, Inc. dba Mindray North America began a recall of Gas Module 3 monitoring spirometer, Part Number: 115-030108-00 Product Usage: The Gas Module 3 is an accessory device designed to monitor airway gases during... on . FDA classified it as Class III and gives the reason as "The firm made multiple changes to the product requiring the submission of a new 510(k) premarket notification". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1242-2018
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Datascope Corp.
Product
Gas Module 3 monitoring spirometer, Part Number: 115-030108-00 Product Usage: The Gas Module 3 is an accessory device designed to monitor airway gases during anesthesia and/or assisted respiration. Monitored gases includes O2, CO2, N2O and anesthetic agents Halothane, Isoflurane, Sevoflurane, Desflurane and Enflurane. It is intended for use in a fixed location such as the operating room or intensive care unit. The Gas Module 3 is compatible with Mindray patient monitors which display the gas analysis results through the use of a proprietary communication protocol.
Product code
BZK Spirometer, Monitoring (W/Wo Alarm)
Reason for recall
The firm made multiple changes to the product requiring the submission of a new 510(k) premarket notification.
Root cause tag
design
FDA root cause
Device Design
Action
Mindray sent an Safety Notification letter dated December 1, 2017 to all customers. The letter identified the affected product, problem and actions to be taken. For question call 201-995-8407.
Quantity in commerce
1719
Distribution
Worldwide Distribution - U.S. Nationwide in the states: AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WY, including Puerto Rico and distributed to 30 foreign consignees.
Product codes and lots
All Gas Modules 3 with Part Number 115-030108-00 purchased between February 14, 2008 and October 2015.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1242-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1242-2018Class IIITerminatedVoluntary: Firm initiated