Z-1223-2018Class II

Vital-Port Vascular Access System, Subcutaneous, implanted, intravascular infusion port and catheter, IP, IP-S

Cook Vandergrift, Inc. / initiated 2017-11-28 / Terminated / root cause: manufacturing process

Cook Vascular Inc. began a recall of Vital-Port Vascular Access System, Subcutaneous, implanted, intravascular infusion port and catheter, IP, IP-S, Product usage: The devices are intended for use... on . FDA classified it as Class II and gives the reason as "Non-coring needle provided with the Cook Vital-Port Vascular Assess System (Vital-Port) may cut or dislodge a core or sliver of material from the Vital-Port septum when the non-coring needle is inserted into the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1223-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Vandergrift, Inc.
Product
Vital-Port Vascular Access System, Subcutaneous, implanted, intravascular infusion port and catheter, IP, IP-S, Product usage: The devices are intended for use in patient therapy requiring long-term vascular access for infusion therapy and/or blood sampling.
Product code
LJT Port & Catheter, Implanted, Subcutaneous, Intravascular
Reason for recall
Non-coring needle provided with the Cook Vital-Port Vascular Assess System (Vital-Port) may cut or dislodge a core or sliver of material from the Vital-Port septum when the non-coring needle is inserted into the Vital-Port. This needle is used on the initial implant of the Vital-Port. Vital-Port products that have been successfully placed in patients are not impacted by this recall.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Removal for all non-expired lots of the Cook Vital-Port Vascular Access System. Written recall communication letters will be mailed via courier to US Consignees.plan to STORE recalled product(s)? (if returned) Recalled products will be returned to global regional distribution warehouses to be quarantined in a marked receiving area of returned goods. 8.9 Provide plan for final DISPOSITION for recalled product(s)? All recalled products will be destroyed at the regional distribution warehouses. A certificate of destruction will be provided upon completion of destruction prior to the completion of the recall. Products remaining in the control of Cook will be assessed for rework to replace the non-coring needle.
Quantity in commerce
31,006 Units
Distribution
The devices have been distributed within the United States and the following countries: United Arab Emirates, Argentina, Austria, Australia, Barbados, Belgium, Bulgaria, Bermuda, Brazil, Canada, Switzerland, Colombia, Costa Rica, Cyprus, Czech Republic, Germany, Denmark, Spain, Finland, Faroe Islands, France, United Kingdom, Guatemala, Hong Kong, Hungary, Indonesia, Ireland, Iceland, Italy, Jamaica, Jordan, Japan, Kenya, South Korea, Kuwait, Lebanon, Latvia, Morocco, Macedonia, Mexico, Netherlands, New Zealand, Oman, Panama, Poland, Portugal, Serbia, Saudi Arabia, Sweden, San Marino, El Salvador, Turkey, Trinidad & Tobago, Ukraine, Uruguay, Venezuela, Virgin Islands, Kosovo, China, and South Africa.
Product codes and lots
IPI, P-S

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1223-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1223-2018Class IITerminatedVoluntary: Firm initiated