Z-1073-2018Class II

PORT-A-CATH II PS/Titan, Tray PAC II. Labeled as: a. VEN LP, PU, 2.6MM, UL, 8.5FR/INTRO (1/EA); b. DL, LP...

Abbott Medical / initiated 2017-12-04 / Terminated / root cause: manufacturing process

Smiths Medical ASD Inc. began a recall of PORT-A-CATH II PS/Titan, Tray PAC II. Labeled as: a. VEN LP, PU, 2.6MM, UL, 8.5FR/INTRO (1/EA); b. DL, LP, PU, 2.2MM, WING-LOCK, 7FR/INTRO (1/EA); c. DL, LP... on . FDA classified it as Class II and gives the reason as "Certain models and lots of PORT A CATH implantable port kits may have been misassembled. The kits may incorrectly contain a 6Fr or 6.5Fr size introducer instead of the correct 8.5Fr size introducer. The Power PORT A...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1073-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
PORT-A-CATH II PS/Titan, Tray PAC II. Labeled as: a. VEN LP, PU, 2.6MM, UL, 8.5FR/INTRO (1/EA); b. DL, LP, PU, 2.2MM, WING-LOCK, 7FR/INTRO (1/EA); c. DL, LP, PU, 3.2MM, WING-LOCK, 10FR/INTRO (1/EA)
Product code
LJT Port & Catheter, Implanted, Subcutaneous, Intravascular
Reason for recall
Certain models and lots of PORT A CATH implantable port kits may have been misassembled. The kits may incorrectly contain a 6Fr or 6.5Fr size introducer instead of the correct 8.5Fr size introducer. The Power PORT A CATH II kit may also contain the incorrect locking collars in addition to the incorrect sized introducer.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Customers were notified via letter on about 12/04/2017. Instructions include to examine current inventory for affected product, collect and isolate any potentially affected product, determine the total number of potentially affected product in possession and complete and return the Urgent Medical Device Recall Response Form, return any affected product on hand, and notify customers if the affected products were further distributed.
Quantity in commerce
1,077 units total
Distribution
Worldwide Distribution. US nationwide including Puerto Rico, Austria, Bahamas, Belgium, Dominican Republic, Netherlands, Spain, and Turks and Caicos.
Product codes and lots
a. Item No. 21-4071-24 (Lot No. 86X046, 87X050);  b. Item No. 21-8066-24 (Lot No. 87X178);  c. Item No. 21-8068-24 (Lot No. 87X179)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1073-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1073-2018Class IITerminatedVoluntary: Firm initiated
Z-1074-2018Class IITerminatedVoluntary: Firm initiated