Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1276-2018Class II

Phadia Prime software, article number 12-4101-00, as used in combination with the Phadia 250 Instrument...

Phadia AB / initiated 2017-11-20 / Terminated / root cause: software

Phadia Ab began a recall of Phadia Prime software, article number 12-4101-00, as used in combination with the Phadia 250 Instrument, article number 12-3900-01 and EliA Assays. This recall... on . FDA classified it as Class II and gives the reason as "We want to inform all Phadia 250 system operators performing EliA Assays not to use the function OK to All in any version of Phadia Prime, up to and including 2.1.4, when rejecting and retesting samples with any EliA...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1276-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Phadia AB
Product
Phadia Prime software, article number 12-4101-00, as used in combination with the Phadia 250 Instrument, article number 12-3900-01 and EliA Assays. This recall is for any Phadia Prime software version up to and including the current version, 2.1.4.
Product code
DHB System, Test, Radioallergosorbent (Rast) Immunological
Reason for recall
We want to inform all Phadia 250 system operators performing EliA Assays not to use the function OK to All in any version of Phadia Prime, up to and including 2.1.4, when rejecting and retesting samples with any EliA assay (the OK function may be used for rejecting single tests and dilution of samples in accordance with product DfU).
Root cause tag
software
FDA root cause
Software design
Action
An urgent recall notification was issued to customers in November of 2017, and informs all Phadia 250 system operators performing EliA Assays not to use the function OK to All in any version of Phadia Prime, up to and including 2.1.4, when rejecting and retesting samples with any EliA assay (the OK function may be used for rejecting single tests and dilution of samples in accordance with product DfU). For further questions, please call (269) 492-1940.
Quantity in commerce
2307
Distribution
US Distribution to the states of :TX, VA, NJ, MA, IN, GA, UT, TN, CA, MS, NC, ME and OR
Product codes and lots
Phadia Prime software 12-4101-00 Phadia 250 12-3900-01  Listed software: EliA dsDNA Well 14-5500-01 EliA U1RNP Well 14-5501-01 EliA Sm Well 14-5502-01 EliA Ro Well 14-5503-01 EliA La Well 14-5504-01 EliA CENP Well 14-5505-01 EliA Scl-70 Well 14-5506-01 EliA Jo-1 Well 14-5507-01 EliA Symphony Well 14-5508-01 EliA RNP70 Well 14-5511-01 EliA GBM Well 14-5514-01 EliA CCP Well 14-5515-01 EliA Celikey IgA Well 14-5517-01 EliA Celikey IgG Well 14-5518-01 EliA Gliadin IgA Well 14-5519-01 EliA Gliadin IgG Well 14-5520-01 EliA Cardiolipin IgA Well 14-5528-01 EliA Cardiolipin IgG Well 14-5529-01 EliA Cardiolipin IgM Well 14-5530-01 EliA B2-Glycoprotein I IgA Well 14-5531-01 EliA B2-Glycoprotein I IgG Well 14-5532-01 EliA B2-Glycoprotein I IgM Well 14-5533-01 EliA PR3S Well 14-5536-01 EliA MPOS Well 14-5537-01 EliA GliadinDP IgA Well 14-5538-01 EliA GliadinDP IgG Well 14-5539-01 EliA RF IgM Well 14-5600-01 EliA RF IgA Well 14-5601-01 EliA SmDP Well 14-5624-01 EliA Scl-70S Well 14-5637-01 EliA anti-TPO Well 14-5641-01 EliA anti-TG Well 14-5642-01 EliA M2 Well 14-5649-01

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1276-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1276-2018Class IITerminatedVoluntary: Firm initiated