Z-1701-2018Class I

MindFrame Capture LP Revascularization Devices: Product Numbers: (a) REF 300010 (b) REF 300011 (c) REF 300012...

Medtronic, Inc. / initiated 2018-02-26 / Terminated / root cause: manufacturing process

Micro Therapeutics Inc, Dba Ev3 Neurovascular began a recall of MindFrame Capture LP Revascularization Devices: Product Numbers: (a) REF 300010 (b) REF 300011 (c) REF 300012 (d) REF 300013 (e) REF 300014 (f) REF 300015 (g)... on . FDA classified it as Class I and gives the reason as "There is a potential issue for all MindFrame Capture devices to partially detach or separate from the delivery wire". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1701-2018
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
MindFrame Capture LP Revascularization Devices: Product Numbers: (a) REF 300010 (b) REF 300011 (c) REF 300012 (d) REF 300013 (e) REF 300014 (f) REF 300015 (g) REF 300016 (h) REF 300017 (i) REF 300018 Intended to restore blood flow by removing thrombus from a large intracranial vessel experiencing ischemic stroke within 8 hours of symptom onset.
Product code
NRY Catheter, Thrombus Retriever
Reason for recall
There is a potential issue for all MindFrame Capture devices to partially detach or separate from the delivery wire.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
The firm initiated their recall via an Urgent Medical Device Recall letter, dated February 26, 2018. The letter identified the affected device, described the issue, product scope, and the actions to be taken. The firm recommended users remove and quarantine all affected and unexpired product in their inventory and return the affected products to Medtronic. Customers may contact their Medtronic representative if alternative product is needed. The Customer Confirmation Form should be completed and faxed to Medtronic at 1-949-434-5020, to the attention of Neurovascular Quality.
Quantity in commerce
1,942 units
Distribution
Worldwide Distribution
Product codes and lots
All lot numbers: Product Number: UDI (a) REF 300010: 00843889031334, 00843889031341, 00847536020207 (b) REF 300011: 00843889031365, 00843889031372, 00847536020214 (c) REF 300012: 00843889031396, 00843889031402, 00847536020221 (d) REF 300013: 00843889031426, 00843889031433, 00847536020238 (e) REF 300014: NOT DISTRIBUTED (f) REF 300015: 00847536020122, 00847536020160 (g) REF 300016: 00847536020139, 00847536020177 (h) REF 300017: 00847536020146, 00847536020184 (i) REF 300018: 00847536020153, 00847536020191

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1701-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1701-2018Class ITerminatedVoluntary: Firm initiated