ARCHITECT DHEA-S Reagent
Abbott Gmbh & Co. KG began a recall of ARCHITECT DHEA-S Reagent Product Usage: It is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of dehydroepiandrosterone... on . FDA classified it as Class II and gives the reason as "Falsely elevated results may be obtained when using the ARCHITECT DHEA-S assay with samples from infants up to 60 days old". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1629-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- ARCHITECT DHEA-S Reagent Product Usage: It is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of dehydroepiandrosterone sulfate (DHEA-S) in human serum and plasma on the ARCHITECT iSystem.
- Product code
- JKC Radioimmunoassay, Dehydroepiandrosterone (Free And Sulfate)
- Reason for recall
- Falsely elevated results may be obtained when using the ARCHITECT DHEA-S assay with samples from infants up to 60 days old.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Customers were notified via letter on about 12/14/2017. Instructions included to immediately discontinue use of the ARCHITECH DHEA-S assay with samples form infants up to 60 days old, to immediately discontinue use of the ARCHITECH DHEA-S reagent insert Expected Values representative data for children up to 10 years old, to review the letter with the Medial Director and follow the laboratory protocol regarding the need for reviewing previously reported patient results, to forward the Product Correction letter to customers if the product was further distributed, and to complete and return the Customer Reply Form. Questions from U.S. customers can be directed to Customer Service at 1-877-4ABBOTT (available 24 hours per day, 7 days a week) and customers outside the U.S. should contact their local area Customer Service.
- Quantity in commerce
- 10,153 units
- Distribution
- Worldwide Distribution in the states of AL, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MA, MN, MO, MS, NC, NJ, NY, OH, OK, PA, PR, SC, SD, TN, TX, UT, VT, WA, WY. Distributed internationally to Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Barbados, Belarus, Belgium, Bosnia and Herzegovi, Botswana, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Egypt, Finland, France, Gaza and Jericho, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Maldives, Mexico, Morocco, Netherlands, Nigeria, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic of Yemen, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Tajikistan, Thailand, Tunisia, Turkey, Uganda, United Arab Emirates, Uganda, United Kingdom, Uruguay .
- Product codes and lots
- List No. 8K2-20 with the following Lot No., Exp. Date, UDI: a. Lot No. 01316K000, Exp. Date 04FEB2018, UDI (01) 00380740017446 (17) 180204 (10) 01316K000 (240) 8K2720; b. Lot No. 01217A000, Exp. Date 13APR2018, UDI (01) 00380740017446 (17) 180413 (10) 01217A000 (240) 8K2720; c. Lot No. 02217D000, Exp. Date 24JUN2018, UDI (01) 00380740017446 (17) 180624 (10) 02217D000 (240) 8K2720; d. Lot No. 04117H000, Exp. Date 25NOV2018, UDI (01) 00380740017446 (17) 181125 (10) 04117H000 (240) 8K2720; List No. 8K27-25 with the following Lot No., Exp. Date, UDI: a. Lot No. 01216K000, Exp. Date 04FEB2018, UDI (01) 00380740017453 (17) 180204 (10) 01216K000 (240) 8K2725; b. Lot No. 01117A000, Exp. Date 13APR2018, UDI (01) 00380740017453 (17) 180413 (10) 01117A000 (240) 8K2725; c. Lot No. 02217C000, Exp. Date 17JUN2018, UDI (01) 00380740017453 (17) 180617 (10) 02217C000 (240) 8K2725; d. Lot No. 02317D000, Exp. Date 24JUN2018, UDI (01) 00380740017453 (17) 180624 (10) 02317D000 (240) 8K2725; e. Lot No. 02417D000, Exp. Date 24JUN2018, UDI (01) 00380740017453 (17) 180624 (10) 02417D000 (240) 8K2725; f. Lot No. 00517F000, Exp. Date 02SEP2018, UDI (01) 00380740017453 (17) 180902 (10) 00517F000 (240) 8K2725; g. Lot No. 02817H000, Exp. Date 21OCT2018, UDI (01) 00380740017453 (17) 181021 (10) 02817H000 (240) 8K2725
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1629-2018 | Class II | Terminated | Voluntary: Firm initiated |