Z-1363-2018Class II

ORTHOFIX, Connector System, SMALL SET SCREW, REF 79-2003

Orthofix SRL / initiated 2018-02-23 / Terminated / root cause: manufacturing process

Orthofix, Inc began a recall of ORTHOFIX, Connector System, SMALL SET SCREW, REF 79-2003 on . FDA classified it as Class II and gives the reason as "Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1363-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Orthofix SRL
Product
ORTHOFIX, Connector System, SMALL SET SCREW, REF 79-2003
Product code
NKG Posterior Cervical Screw System
Reason for recall
Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm disseminated the recall notices on 02/23/2018 by mail. The firm requested the return of the screws.
Quantity in commerce
397 units
Distribution
US
Product codes and lots
UDI 18257200125371, Lot Numbers: O02, O03 and O04

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1363-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1363-2018Class IITerminatedVoluntary: Firm initiated
Z-1364-2018Class IITerminatedVoluntary: Firm initiated