ORTHOFIX, Connector System, SMALL SET SCREW, REF 79-2003
Orthofix, Inc began a recall of ORTHOFIX, Connector System, SMALL SET SCREW, REF 79-2003 on . FDA classified it as Class II and gives the reason as "Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1363-2018
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Orthofix SRL
- Product
- ORTHOFIX, Connector System, SMALL SET SCREW, REF 79-2003
- Product code
- NKG Posterior Cervical Screw System
- Reason for recall
- Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm disseminated the recall notices on 02/23/2018 by mail. The firm requested the return of the screws.
- Quantity in commerce
- 397 units
- Distribution
- US
- Product codes and lots
- UDI 18257200125371, Lot Numbers: O02, O03 and O04
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1363-2018 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1364-2018 | Class II | Terminated | Voluntary: Firm initiated |