510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1199-2018Class II

DATEX-OHMEDA AISYS CS2 ALADIN2 CASSETTE Cassettes are intended to be used to deliver anesthetic agent

GE Healthcare / initiated 2017-10-04 / Terminated / root cause: manufacturing process

GE Healthcare, LLC began a recall of DATEX-OHMEDA AISYS CS2 ALADIN2 CASSETTE Cassettes are intended to be used to deliver anesthetic agent on . FDA classified it as Class II and gives the reason as "Non-Conforming devices were shipped from the service center without proper evaluation and therefore may be nonconforming". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1199-2018
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
DATEX-OHMEDA AISYS CS2 ALADIN2 CASSETTE Cassettes are intended to be used to deliver anesthetic agent.
Product code
BSZ Gas-Machine, Anesthesia
Reason for recall
Non-Conforming devices were shipped from the service center without proper evaluation and therefore may be nonconforming.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm, GE healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated October 4th, 2017 to all their customers. The letter described the product, problem and actions to be taken. The customers were instructed to contact GE Service to obtain an RMA number. GE Healthcare has arranged with Piramal Inc. to replace the affected products at no cost. If you have any questions or concern regarding this notification please contact GE healthcare Service at 1-800-437-1171 or your local representative.
Quantity in commerce
2
Distribution
US Distribution to states of: AZ, IL NC, NH and PA.
Product codes and lots
Model 1100-9034-000 Mfg. Lot or Serial # BGBQ00675 and BGBQ01187

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1199-2018
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1198-2018Class IITerminatedVoluntary: Firm initiated
Z-1199-2018Class IITerminatedVoluntary: Firm initiated